Visx Star S3
|
 |
The VISX STAR platform's heritage of dependability and performance is well known. The S3 excimer laser system has a faster chair, automatic alignment for both eyes, horizontal alignment capability, multiple treatments and offline programming, and the following exciting features: Variable Spot Beam Technology: Variable-shaped beams ranging in size from 0.65 to 6.5 millimeters are scanned* and offset to allow personalized refractive treatments.
|
Exclusive SmartBeam technology adjusts beam size according to treatment, minimizing corneal tissue removal. *The VISX laser utilizes the VSS technology in creating a blend zone in myopic patients receiving1.0 diopter of vision correction. ActiveTrak 3-D Active Eye Tracking: The only laser system that can capture all three dimensions of intra-operative eye movements without the requirement of pupillary dilation. Larger Ablation Zones: VSS (Variable Spot Scanning) technology adds a blend zone that increases the ablation zone by 51%, increasing the overall diameter of the ablation to 8 millimeters.
|
|
- 10 ft by 13 ft (3.1 m by 4 m) Door opening: 36 in (92 cm)
- 12 ft by 15 ft (3.6 m by 4.5 m) or larger
- Ambient temperature: 60 to 80° F (15 to 27° C)
- Humidity requirements: Relative humidity between 35%-65% (non-condensing)
- Barometric pressure between 11 to 16 psi (1.6 to 2.3 kPa, 12 to17.35 mmHg)
- Room should be kept clean; dust-/particle-free by tiling floor; do not install carpet; avoid ceiling tiles that shed particles ; no volatile cleaners
- Height: . . . . . . 59 in/149 cm maximum
- Width: . . . . . . . 50 in/127 cm overall
- Length: . . . . . . 80 in/203 cm
- Weight: . . . . . . Main console: 1600 lb/726 kg
- Patient chair: . . 400 lb/181 kg
- Patient chair: . . Perpendicular to system at treatment end
Follow all local codes and standards. Electrical conduit to the room must carry:
- 220 VAC single phase, plus ground, 60 Hz (U.S. specifications)
- 230 VAC single phase, plus ground, 50 Hz (European specifications)
- 30 amp service
Note: Power must be greater than 199 VAC and less than 245 VAC |
|