1. Treatment Area (Technolas 217z and Hansatome Excellus) |
Heating, Ventilation, Air Conditioning |
Room temp: |
18°-24° C / 64 - 75 F respectively |
Humidity: |
40% (ideal); 30 - 50% (range) |
Air circulation capacity: |
12 air exchanges/h (room volume in _ 5min) |
Contaminants: |
Minimize dust; no smoking; no volatile cleaning; no agents (alcohol, ketone, etc.) |
Voltage:
208/230VC ( _ 10%), 60 Hz
Amps:
16A (without UPS); 50A (with UPS supplied by B&L)
Cable length:
8' 2" (2,5 m) |
Weight capacity > 0.435lb/sqin, > 62.65lb/sqft, (> 3000N/sqm.) Standard reinforced concrete floor; flatness _ 1/4 in/yard ( _ 6mm/m)
|
Paint, hard vinyl covering, no carpet.
|
Low particulate ceiling tile free of odor, vapor or other impurities which can be discharged. It must be possible to darken the room completely, e.g. blinds, etc. (Ventilate room at least two weeks after painting/flooring) |